|
Maintenance Tips
HIGH SPEED HANDPIECE MAINTENANCE GUIDELINES
AFTER EACH PATIENT |
- Remove handpiece from unit tubing, clean with alcohol (Do Not use harsh cleaning agents, Do Not place in ultra sonic unit).
- With bur inserted, place lubricant (2 or 3 drops) into air drive port, - at the same time, rotate the bur between your fingers forward/backward 1/4 turns for a few seconds.
- Re-connect handpiece to unit tubing, place an autoclave bag over the handpiece & run 30 seconds to eliminate excess oil from the turbine. (This procedure is primarily to clean the handpiece not lubricate it).
- Remove from unit tubing, and autoclave.
- After autoclaving, allow to cool gradually, about 30 minutes, to room temperature.
- With bur in handpiece place lubricant (2 or 3 drops) into air drive port, - at the same time, rotate the bur between your fingers forward/backward 1/4 turns for a few seconds.
- Connect handpiece to unit tubing & run for 30 seconds to eliminate excess oil from the turbine (this procedure is to lubricate the handpiece).
NOTE: A liquid lubricant (oil) seems to work best.
>>>>>>>SET AIR PRESSURE TO MANUFACTURE'S SPECIFICATIONS<<<<<<<
|
LUBRICATING CONTRA ANGLES AND HEADS
AFTER EACH PROCEDURE
- Disassemble head from Contra Angle as shown in diagram.
- Wipe off parts with alcohol, NEVER place any parts in an ultra sonic unit!
- Place a lubricant (2 or 3 drops) into all parts noted by the arrow.
- Reassemble, the Contra Angle and place it on a Slow speed motor and run for 30 seconds (this is for cleaning purposes).
- Autoclave and let cool.
- Repeat steps 1 thru 4 (this is for lubricating purposes).
|
|
|
|
|